EU (Engineering Unit Settings)
.................................. This menu is used to scale the measured values and convert them to
Function (EU) ...................... Sets the EU function to Off or On.
Lower - EU - Upper Unit ..... Sets the EU function's conversion value and unit. If the ENTER key is
other units.
pressed here, the following window is displayed.
(a) Meas. Value
(e) Select
(d) Unit
If the following message appears, follow the instructions to reduce the
number of digits to be output by one.
Out of input range
[ENTER] Apply
CHECKPOINT
The Scaling operation is calculated using a ratio of the Meas. Value or
EU Value settings. If a ratio value that the GL500 cannot process is
specified, the message above appears.
(a) Meas. Value
Specifies the numeric value to be converted. Set two points, the
Upper and Lower parameters.
(b) EU Value
Specifies output after conversion. Set two points, the Upper and
Lower parameters.
(c) Dec pt
This parameter specifies the decimal point position of the values
specified for EU Value.
(d) Unit
Specifies the converted unit. The unit can be specified as a user-
defined character string consisting of alphanumeric text. It can also
be specified by using the Select setting to select a unit.
(e) Select
Selects the type of engineering unit.
(f) Choose
Selects the converted unit. The unit displayed here is the unit for
the type of engineering unit selected by the Select setting.
To specify a unit that is not displayed here, specify a user-defined
character string as the Unit setting. Moreover, the setting specified
here is displayed as the Unit setting.
Setting Example
Upper Value
Specified Value
+2.5000
EU Value
+10.000
+5 V
CH.1 10V
–5 V
(b) EU Value
(c) Dec pt
Lower Value
–2.5000
–10.000
+20.00 rpm
CH.1 Scaling 1
–20.00 rpm
Settings and Measurement
(f) Choose
3-13