Connecting to the electrical outlet
CAUTION
NOTE:
CAUTION
CAUTION
LOGIQ V2/LOGIQ V1
Direction 5610736-100 Rev. 9
To connect the system to the electrical supply:
1. Ensure that the wall outlet is of the appropriate type.
2. Unwrap the power cable. Make sure to allow sufficient slack
in the cable so that the plug is not pulled out of the wall if the
system is moved slightly.
Use the appropriate power cord provided by or designated
by GE.
3. Attach the power plug to the system.
4. Connect the power cable to the adapter, if it is not
connected.
5. Push the power plug securely into the wall outlet.
Do not use an extension cord or adapter plug.
Disconnect the plug from the wall outlet in case an emergency
should occur. Ensure easy access to the power outlet.
To avoid leakage current above safety limits as prescribed by
IEC 60601-1 and to ensure continuity of protective earth. Only
connect LOGIQ V2/LOGIQ V1 and mains-operated
accessories to the appropriate wall outlet. DO NOT connect
them to a single or multiple socket outlets, an extension cord,
power strip or an adapter plug.
Figure 1-31. Connect the system to the electrical supply
–
User Guide
System Start-Up
1-47