Performing an Exam
TAMAX and TAMEAN (continued)
Auto Trace
2-58
The value measured depends upon the Vol Flow Method preset.
The two selections available are: Peak (TAMAX) and Mean
(TAMEAN).
To auto trace TAMAX:
1. Press Measure key to enter into measure mode, if
necessary.
Adjust Caliper Change in the Primary Menu to S/D Trace.
In the Utility -> Measure - Advanced preset menu set Trace
preset to Auto.
2. To position the caliper at the trace start point in the Doppler
spectrum, move the Trackball.
3. To fix the start point, press Set.
4. To position the vertical caliper at the end point, move the
Trackball.
5. To complete the measurement, press Set. The system
automatically fixes both calipers and traces the maximum
value between the two points. The system displays this
value in the Results Window.