Performing an Exam
M-Mode Measurements
NOTE:
Tissue depth
2-60
Basic measurements that can be taken in the M-Mode portion of
the display are:
•
Tissue Depth (Distance)
•
Time Interval
•
Time Interval and Velocity
The following instructions assume that you do the following:
1. In the B-Mode part of the display, scan the anatomy you
want to measure.
2. Go to the M-Mode part of the display.
3. Press Freeze.
Tissue depth measurement in M-Mode functions the same as
distance measurement in B-Mode. It measures the vertical
distance between calipers.
1. Press Measure key to enter into measure mode, if
necessary.
Adjust Caliper Change in the Primary Menu to MM
Distance.
An active caliper with a vertical and horizontal dotted line
displays.
2. To position the active caliper at the most anterior point you
want to measure, move the Trackball.
3. To fix the start point, press Set.
The system fixes the first caliper and displays a second
active caliper.
4. To position the second caliper at the most posterior point
you want to measure, move the Trackball.
5. To complete the measurement, press Set.
The system displays the vertical distance between the two
points in the Results Window.
LOGIQ V2/LOGIQ V1
Direction 5610736-100 Rev. 9
–
User Guide