Discharge/Charge Cycle - GE LOGIQ V2 User Manual

Hide thumbs Also See for LOGIQ V2:
Table of Contents

Advertisement

Discharge/Charge Cycle

NOTE:
CAUTION
LOGIQ V2/LOGIQ V1
Direction 5610736-100 Rev. 9
When the battery is stored for three months or more, the
customer should perform one full discharge/charge cycle.
A full discharge/charge cycle means the system is turned on
using battery power until the battery loses its charge completely
and the system shuts down. Plug the LOGIQ V2/LOGIQ V1 in
until the battery is fully charged as indicated by a green LCD
light.
Upon receipt of the LOGIQ V2/LOGIQ V1 and before first time
usage, it is highly recommended that the customer perform one
full discharge/charge cycle.
If the battery has not been used for >2 months, the customer is
recommended to perform one full discharge/charge cycle. It is
also recommended to store the battery in a shady and cool area
with FCC (full current capacity).
One Full Discharge/Charge Cycle Process:
1. Full discharge of battery to let the LOGIQ V2/LOGIQ V1
automatically shut down.
2. Charge the LOGIQ V2/LOGIQ V1 to 100% FCC (full current
capacity).
3. Discharge of LOGIQ V2/LOGIQ V1 for complete shut down
(takes one hour for discharge).
When storing packs for more than 6 months, charge the pack at
least once during the 6 month time frame to prevent leakage
and deterioration in performance.
Use only GE recognized batteries.