Related Hazards (continued)
Smoke &
Fire Hazard
WARNING
WARNING
Explosion
Hazard
Explosion
Hazard
CAUTION
LOGIQ V2/LOGIQ V1
Direction 5610736-100 Rev. 9
The system must be supplied from an adequately rated
electrical circuit. The capacity of the supply circuit must be as
specified.
Only approved and recommended peripherals and accessories
should be used.
All peripherals and accessories must be securely mounted to
the LOGIQ V2/LOGIQ V1.
The LOGIQ V2/LOGIQ V1 is not intended to be used as a data
storage device; backup of the Patient and Image Database is
your institution's responsibility. GE is NOT responsible for any
lost patient information or for lost images.
Risk of explosion if used in the presence of flammable
anesthetics.
Never operate the equipment in the presence of flammable or
explosive liquids, vapors or gases. Malfunctions in the unit, or
sparks generated by fan motors, can electrically ignite these
substances. Operators should be aware of the following points
to prevent such explosion hazards.
If flammable substances are detected in the environment,
do not plug in or turn on the system.
If flammable substances are detected after the system has
been turned on, do not attempt to turn off the unit, or to
unplug it.
If flammable substances are detected, evacuate and
ventilate the area before turning off the unit.
Do not unpack the LOGIQ V2/LOGIQ V1. This must be
performed by qualified service personnel only.
User Guide
Safety Precautions
4-11