Resetting the Test Sequence Number
The test sequence numbers run from 0001 to 9999. You can reset the
test sequence number to start at 0001 for the next test.
To reset the test sequence number, perform the following steps:
1. On the Select Ready screen, select Instrument Set Up.
2. On the Choose Settings screen, select Test sequence number.
The Sequence Number screen shows the next test number.
3. Select Reset to 0001.
4. Select Done twice to return to the Select Ready screen.
Changing the Instrument Settings
The instrument settings control the way the system displays
information and operates.
The instrument settings include the following options:
•
Results Format
•
System Settings
•
Display Contrast
•
Connectivity
•
Urinalysis Test Settings
•
Authorized Operator
•
Printer Settings
•
QC Settings
•
Software Update
To change the instrument settings, perform the following steps:
1. On the Select Ready screen, select Instrument Set Up.
2. On the Choose Settings screen, select Instrument Settings.
3. On the Instrument Settings screen, select an option.
4. Select Next or Previous to move forward and backward through
the screens.
5. When you finish, select Done.