Note An ID band is a white or colored area near the handle of a
Siemens urinalysis strip. CLINITEK Microalbumin and Multistix PRO
urinalysis strips contain a colored ID band. The analyzer reads the ID
band to identify the strip type. Therefore, you do not need to select the
strip type from a menu.
To prepare a urinalysis strip Quick Test, perform the following steps:
Note If you use a reagent strip that has 4 or fewer test pads, such as
Uristix 4 reagent strips, use a short test table insert. You need to order
the short test table insert separately from the analyzer (see Appendix
C, Orderable Supplies).
1. On the Select Ready screen, select Strip Test.
2. If you enabled lot information with Instrument Set Up, enter the
strip lot number and expiration date, as follows. Otherwise, go to
step 3.
To use the last strip number and begin the test, select Use Last
Lot.
To enter new strip data, select Enter new lot and expiration.
Enter the strip lot number and select Enter. Use the arrow keys
to enter the strip expiration date and select Enter.
3. Make sure the reagent strip holder faces upward in the test table
insert.
4. Have the urinalysis strip and paper towel ready.

Running a Urinalysis Strip Quick Test

When you run a urinalysis strip Quick Test, the analyzer calibrates and
then analyzes the strip.
To run a urinalysis strip Quick Test, perform the following steps:
Note After you select START, you have 8 seconds to dip the
reagent strip in the urine sample and place the strip in the test
table channel.
30
CLINITEK Status+ Analyzer Operator's Guide