Summary of Contents for Siemens CALOMAT 6

  • Page 3 CALOMAT 6 Thermal Conductivity Gas Analyzers SIL Safety Manual Supplement to instruction manual CALOMAT 6 CALOMAT 6F 7MB2511, 7MB2517 CALOMAT 6E 7MB2521, 7MB2527 Ordering Number: A5E00729184 07/2010...
  • Page 4 Trademarks All names identified by ® are registered trademarks of the Siemens AG. The remaining trademarks in this publication may be trademarks whose use by third parties for their own purposes could violate the rights of the owner.
  • Page 5: Table Of Contents

    Behavior in case of faults ..................11 Maintenance / Checking ..................11 Safety characteristics.................... 13 Appendix ........................14 SIL Declaration of Conformity ................14 exida Test Report (extract)................15 List of Abbreviations/Acronyms ..................17 Abbreviations ....................17 Glossary........................18 Safety Manual CALOMAT 6 Gas Analyzers 07/2010, A5E00729184-02...
  • Page 6: Introduction

    It is aimed at system planners, constructors, service and maintenance engineers and personnel who will commission the device. Required documentation This document deals with the CALOMAT 6 gas analyzer exclusively as part of a safety function. This document only applies in conjunction with the following documentation: No.
  • Page 7: Change History

    The contents of these instructions shall not become part of or modify any prior or existing agreement, commitment or legal relationship. All obligations on the part of Siemens AG are contained in the respective sales contract which also contains the complete and solely applicable warranty conditions. Any statements contained herein do not create new warranties or modify the existing warranty.
  • Page 8: General Safety Instructions

    The analyzer logic unit/control system is a final controlling element combine to form a safety-instrumented system, which executes a safety function. Note This document deals with the CALOMAT 6 exclusively as part of a safety function. SIL Safety Manual CALOMAT 6 Gas Analyzers A5E00729184-02, 07/2010...
  • Page 9 Function The measuring principle is based on the different thermal conductivity of gases. The CALOMAT 6 operates with a micro-mechanically manufactured Si chip whose measuring diaphragm contains thin-film resistors. The resistors are regulated at a constant temperature. A current is required to achieve this, and its magnitude assumes a particular value depending on the thermal conductivity of the sample gas.
  • Page 10: Safety Integrity Level (Sil)

    B systems include sensors and positioners actuators with complex components, e.g. microprocessors (see also IEC 61508, Section 2). <60% Not allowed SIL1 SIL2 60 to 90% SIL1 SIL2 SIL3 90 to 99% SIL2 SIL3 SIL4 >99% SIL3 SIL4 SIL4 SIL Safety Manual CALOMAT 6 Gas Analyzers A5E00729184-02, 07/2010...
  • Page 11: Device-Specific Safety Instructions

    The CALOMAT 6 is usable in safety applications to monitor limits. Safety function The CALOMAT 6 is mainly used for user defined threshold monitoring. Only the analog output 4.20 mA (NAMUR) was calculated as safety function. The danger- ous failure is a deviation of the output current of ±5% full span.
  • Page 12: Settings

    Checking the safety function after installation After installation of the CALOMAT 6 a safety function test has to be carried out (see chapter 4 “Maintenance“ and 5 “Operation“ of the instruction manual CALOMAT 6E/F).
  • Page 13: Behavior In Case Of Faults

    Maintenance / Checking Checking function We recommend that the functioning of the CALOMAT 6 is checked at regular intervals of one year. Check at least the following: Test the basic functionality of the CALOMAT 6 as described in the Instruction Manual /1/.
  • Page 14 Generate or simulate an alarm condition to force the Gas Analyzer CALOMAT 6 to go to the high alarm current output and verify that the analog current reaches that value. Generate or simulate an alarm condition to force the Gas Analyzer CALOMAT 6 to go to the low alarm current output and verify that the analog current reaches that value.
  • Page 15: Safety Characteristics

    (see chapter Appendix). These values apply under the following conditions:  The CALOMAT 6 is only used in safety-related systems with a low demand mode for the safety function.  The safety-related parameters/settings (see “Settings” section) have been entered by local operation and checked before commencing safety- instrumented operation.
  • Page 16: Appendix

    Siemens AG Appendix SIL Declaration of Conformity SIL Safety Manual CALOMAT 6 Gas Analyzers A5E00729184-02, 07/2010...
  • Page 17: Exida Test Report (Extract)

    Siemens AG A.2 exida Test Report (extract) Safety Manual CALOMAT 6 Gas Analyzers 07/2010, A5E00729184-02...
  • Page 18 Siemens AG SIL Safety Manual CALOMAT 6 Gas Analyzers A5E00729184-02, 07/2010...
  • Page 19: List Of Abbreviations/Acronyms

    In a 1oo1 architecture, there is only one pressure sensor. Safety Manual CALOMAT 6 Gas Analyzers 07/2010, A5E00729184-02...
  • Page 20: Glossary

    The achievable SIL is determined by the following safety characteristics:  Average probability of dangerous failure of a safety function in case of demand (PFD  Hardware fault tolerance (HFT)  Safe failure fraction (SFF) SIL Safety Manual CALOMAT 6 Gas Analyzers A5E00729184-02, 07/2010...