7. Rinse the test table and the table insert thoroughly with water.
CAUTION
Rinse away all the solution residue, as any remaining solution
might affect the reagent pad chemistries.
8. Dry the test table and the table insert thoroughly with a soft cloth,
except for the white calibration bar.
9. Insert the test table and the table insert in the analyzer, as
described in Weekly Cleaning of the Test Table and Test Table
Insert‚ page 57.

Cleaning the Outside of the Analyzer

Always keep the outside of the CLINITEK Status+ analyzer clean and
free of dust.
BIOHAZARD
Wear personal protective equipment. Use universal
precautions. For recommended precautions when you work
with biohazardous materials, see Appendix A, Safety
Information.
To clean the outside of the analyzer, perform the following steps:
1. Power off the analyzer by pressing the on/off button for 2 seconds.
2. Wipe the outside (including the display) with a damp (not wet)
cloth and a mild detergent.
CAUTION
Do not use any type of solvent, oil, grease, silicone spray, or
lubrication on the analyzer.
Do not spray glass cleaner directly onto the screen.
Do not use laboratory wipes, such as Kimwipes, because they
might scratch the screen.
Prevent liquid from entering inside the printer compartment.
You could damage the analyzer or the printer.