Running Diagnostics

You can run the following diagnostics on the analyzer:
Display
Touch Screen
Printer
Test Table
Light Source
Electronics
Check cassette
CAUTION
Do not run the Check Cassette diagnostics on your own. Run
the Check Cassette diagnostic tests only when your local
technical support provider or distributor asks you to do so.
The representative will lead you through the test procedure.
For local technical support providers and distributors, see
Appendix B, Support Information.
To run the diagnostics, perform the following steps:
1. On the Select Ready screen, select Instrument Set Up.
2. On the Choose Settings screen, select Instrument Settings.
3. On the Instrument Settings screen, select Diagnostics.
4. On the Select Diagnostics Test screen, select a diagnostic test
5. Select Select.
6. Read the onscreen instructions.
7. Select Run Test.
8. Select Done twice.

Viewing the System Information

You can view the following system information:
Serial number
Software version