Siemens CLINITEK Status+ Operator's Manual page 31

Table of Contents

Advertisement

1. Select START.
The Prepare Test screen displays steps on how to perform the test.
A timer displays the amount of time remaining that you have to
complete the task.
Note To display the strip testing steps on the screen, select Help.
2. Dip the reagent strip in the urine sample and wet all the pads.
The ID band provides auto-strip identification to ensure that the
analyzer reports the correct strip configuration when you perform
a urinalysis test. The analyzer also performs a strip integrity check
for exposure to humidity.
Note Be sure to use the proper dipping technique.
3. Immediately remove the strip from the urine.
4. Drag the edge of the strip against the side of the sample container
as you remove it.
5. Blot the edge of the strip on a paper towel to remove the excess
urine.
6. Place the reagent strip in the test table channel with the test pads
facing up.
7. Slide or push the strip to the end of the channel. Do not touch the
pads on the strip.
After the 8-second countdown ends, the analyzer pulls in the
test table and strip, and then calibrates.
Note Each time you run a test, the analyzer calibrates.
CAUTION
Do not push or pull the test table because the calibration
might fail or the movement might cause might cause table
positioning errors.
Do not move or bump the table while the analyzer calibrates.
The calibration might fail.
After the calibration finishes, the analyzer starts analyzing the
strip, and the Analyzing screen displays.
CLINITEK Status+ Analyzer Operator's Guide
31