Note By default, the custom header is disabled. However, if you
enter information for the header, the system enables the custom
header printout option.
Select Enter Line 1 and enter the information for the first line
of the custom header.
Select Enter Line 2 and enter the information for the second
line of the custom header.
Note Enter up to 24 alphanumeric characters for each header line.
8. Select Next. On the Printer Settings Printer options 4 of 4
screen, select or clear any of the following options:
Note To use an external printer, connect and enable the Status
connector.
To print to an external printer, select External printer.
Note If you select External printer, sample interference notes
print automatically.
To print to the internal printer, select Internal printer
(default).
If you select Internal printer and want to print sample
interference notes, select Enabled for the Print notes on
internal printer option.
To disable printing sample interference notes, select Disabled.
9. Select Done 3 times to return to the Select Ready screen.

Changing the Quality Control Settings

The Quality Control settings display but they are disabled. If you run
the CLINITEK Status+ analyzer with a CLINITEK Status connector, the
Quality Control settings are available. For instructions on how to set up
the Quality Control settings, see the CLINITEK Status Connect System
Operator's Guide.

Restoring the Default Settings

You can restore the original settings (see Default Settings‚ page 85) for
the system.
CLINITEK Status+ Analyzer Operator's Guide