Siemens CLINITEK Status+ Operator's Manual page 71

Table of Contents

Advertisement

Error
Error Message
Code
E53
Strip Test selected
but cassette
detected
E54
Cassette Test
selected but strip
detected
E56
Incorrect size test
table
E57
Missing strip or
cassette
E58
Misplaced strip
CLINITEK Status+ Analyzer Operator's Guide
Action
Repeat the test using the Cassette
Test routine (see Section 2,
Operations, Performing an hCG
Cassette Quick Test‚ page 35).
Repeat the test using the Strip Test
routine (see Section 2, Operations,
Performing a Urinalysis Strip Quick
Test‚ page 29).
Repeat the test using the correct test
table (see Section 2, Operations,
Performing a Urinalysis Strip Quick
Test‚ page 29).
Repeat the test and ensure that you
correctly position the strip or
cassette on the test table (see
Section 2, Operations, Performing a
Urinalysis Strip Quick Test‚ page 29
or Performing an hCG Cassette Quick
Test‚ page 35).
1. Repeat the test and ensure that
you correctly position the strip
on the test table (
Operations
Performing a
,
Urinalysis Strip Quick Test‚
page 29).
2. If the error remains, examine the
test table insert to ensure that
the small, white line located
near the tip of the strip (on the
strip side of the insert) is present
and not damaged.
3. If this line is damaged, contact
your local technical support
provider or distributor.
see Section 2,
71