Waters 2707 Operator's Manual page 57

Hide thumbs Also See for 2707:
Table of Contents

Advertisement

4.
Loosen the nut that connects the tubing to port 4 of the injection valve.
5.
Pull the sample needle out of its fitting by the tubing.
6.
Install a new needle assembly.
Recommendation:
7.
Using the needle connection nut, tighten the needle assembly.
8.
Connect the other end of the needle connection tubing to port 4 of the
injection valve.
Recommendation:
9.
In the new Replace Needle window, click Reset.
Make sure the air seal encircles the needle.
To avoid blocking the tubing, finger tighten only.
Nut connecting tubing
to port 4
Tubing
Needle connection nut
Chrome locking nut
Air needle
Replacing the sample needle
3-7