Set The Pressure Module To Zero; Example Procedure: Measure Pressure - GE Druck DPI 620 User Manual

Advanced modular calibrator
Hide thumbs Also See for Druck DPI 620:
Table of Contents

Advertisement

4.4.3 Set the pressure module to zero

TAP the window
(Maximise)
1
See the new reading

4.4.4 Example procedure: Measure pressure

1
2
[EN] English - K0449
Use this option to write a new zero pressure value to the
pressure module you are using.
The sensor adjustment is permitted if it obeys this condition:
Adjustment d 10% FS positive pressure value
Note: To make a temporary adjustment for zero, you can use
the Tare function; see
Select Settings
2
5
Go back to the Home window.
3
P1
1. Assemble the pressure indicator with the correct PM 620
modules; see
(for the Sensor)
Section 2.11.1
Select Zero
Accept the adjustment value
3
p1
p2
P2
Section
4.2.1.
Pressure indicator operation (MC 620) 4-5
Issue 1
4
p1
p2

Hide quick links: