Run A Procedure - GE Druck DPI 620 User Manual

Advanced modular calibrator
Hide thumbs Also See for Druck DPI 620:
Table of Contents

Advertisement

Issue 1
Input and Reference
options
Continued
from step 8

7.3 Run a procedure

7-2 Documenting functions
Channel type
Scaling
B
(Input only): Values for Reference High and Low
and Input High and Low. This sets the scale for the Analysis
function.
Error Type
C
Linearity
D
(Input only): Linear or Square Root
Tolerance
E
pass/fail calculation
When the Input and Reference options are set, start the
analysis.
Start the
Set applicable values. When the Analysis
Analysis
work is complete, tap on exit.
9
The Intecal calibration Procedure contains all the values to
calibrate a device under test (test points, ramp time). You can
use the same calibration procedure for all the applicable
devices under test.
To use the Run Procedure function, you need these items:
a copy of the Intecal Calibration Software. This is available
from our website at: www.gesensinginspection.com
There is a 30 day free trial.
IO620-USB-RS232 (Intecal version 5 only). An adaptor to
connect the DPI 620 calibrator to the RS232 interface on
your computer.
a DPI 620 calibrator device driver: available as a download
from our website.
: Input or Reference
A
(Input only): % Span or % Rdg (Reading)
(Input only): an applicable value for the
10
in specified
tolerance limit
not in specified
tolerance limit
K0449 - [EN] English