3.2.8 Example procedure: Measure or source frequency signals

A
Measure frequency on
channel 1 (CH1)
Range: 0 to 50 kHz
Trigger level: 2.5 V
B
Source frequency on
channel 1 (CH1)
Range: 0 to 50 kHz
Waveform: Triangle
Amplitude: 12.0 V
Automation: Nudge
(Section
2.11.2)
Example A
Example B
3-6 Electrical and IDOS operations
1. Set the applicable software options; see
(Procedure
overview).
2. Complete the electrical connections and continue with the
measure or source operation.
3. Source only (Automation): Set the applicable output value;
see
Section
2.11.2.
These examples (A and B) show Channel 1 (CH1) set up to
measure or source a frequency. This includes Hz, kHz and
counts (cpm or cph).
The selection of units sets the available range, for example:
Hz = 0 to 1000 Hz
1
3
1
3
4
1. Set the applicable software options; see
(Procedure
overview).
2. Complete the electrical connections.
3. If necessary change the Trigger Level (Settings) and
continue with the measure operation.
Values to set:
Mode (Automatic/Manual);
Manual Level (trigger level value)
1. Set the applicable software options; see
(Procedure
overview).
Section 3.2.1
kHz = 0 to 50 kHz
2
2
Section 3.2.1
Section 3.2.1