GE Druck DPI 620 User Manual page 55

Advanced modular calibrator
Hide thumbs Also See for Druck DPI 620:
Table of Contents

Advertisement

4.4 Measure
pressure
4.4.1 Procedure overview
3
2
1
TAP
A
3
B
C
D
1
[EN] English - K0449
Step
Procedure
4
4.
Re-attach the adaptor to the MC 620 carrier and tighten
it until it is hand tight only.
When the pressure indicator assembly is complete
(Section
4.2.1), use the menus to set up the necessary
operations.
To use the pressure indicator, complete these procedures:
Set the calibrator functions you want to use on the display;
see
Section
a. Pressure function (P1 and/or P2): item
b. CH1: Channel 1 electrical function (measure or source)
c. CH2: Channel 2 electrical function (measure or source)
d. other functions (maximum: 6 functions)
If necessary, change the Units for the function: item
see
Section 2.10.5
If necessary, set a Utility for the function: item
Section 2.10.6
a. Max/Min/Avg
b. Switch Test
c. Leak Test; see
If necessary, change the Settings for the pressure function:
item
D
a. Process (Tare, Alarm, Filter, Flow, Scaling);
see
Section 2.11.1
b. Leak Test (Only when the Utility is set); see
c. Zero. We recommend that you zero a gauge sensor
before you use it; see
2.10.3. This includes:
Section 4.4.2
Section 4.4.3
Pressure indicator operation (MC 620) 4-3
Issue 1
;
A
;
B
C
; see
Section 4.4.2