Example Procedure: Measure Ac Voltage (Ch1) With The Ac Probe - GE Druck DPI 620 User Manual

Advanced modular calibrator
Hide thumbs Also See for Druck DPI 620:
Table of Contents

Advertisement

Issue 1
A
Measure AC volts on
channel 1 (CH1)
Range: 0 to 20 V(rms)
B
Measure AC millivolts on
channel 1 (CH1)
Range: 0 to 2000 mV(rms)

3.2.5 Example procedure: Measure AC voltage (CH1) with the AC probe

1
3-4 Electrical and IDOS operations
1
1
1. Set the applicable software options; see
(Procedure
overview).
2. Complete the electrical connections and continue with the
measure operation.
To prevent electrical shocks, use only the GE specified
AC probe (Part: IO620-AC) to measure AC voltages that
are more than 20 Vrms (maximum: 300 Vrms). Attach it
to the specified connections only.
This example shows Channel 1 (CH1) set up to measure an
AC voltage with the AC probe (maximum: 300 Vrms).
2
AC
1. Set the applicable software options for the AC Probe; see
Section 3.2.1 (Procedure
2. Complete the electrical connections.
Red: V/Hz connector; Black: COM connector
Then continue with the measure operation.
Maximum: 0 to 20 Vrms
2
AC
Section 3.2.1
WARNING
Maximum: 300 Vrms
overview).
K0449 - [EN] English

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Druck mc 620Druck pm 620

Table of Contents