2.5.3 Charge the battery

2.5.4 Charge times

2.5.5 Operating time

Power save options
2.6 The DC power
supply
Status bar:
Charging symbols
[EN] English - K0449
You can charge the battery in the instrument
you can use the optional external battery charger; see
Section 1.4
(Accessories).
Charge method
DC power supply
External battery charger
USB mini Type B connector
Note: You can use the instrument while the DC charger is
charging the battery but this increases the charge time.
Operation
Continuous operation (measure)
Continuous operation (measure
and source with loop power on)
These are typical operating times for a new, fully charged
Li-polymer battery pack with these settings:
Backlight Intensity set to 80% (Default: 80%)
Backlight Timeout set to 2 hours (Default: 2 minutes)
To get the best battery duration, set a low value for the
Backlight Intensity (40%) and a short Timeout; see
Section 2.10.1 (Procedure to set the basic operations:)
To prevent an explosion or fire, use only the GE
specified battery, power supply and battery charger.
To prevent battery leakage or heat generation, only use
the battery charger and power supply in the
temperature range 0 to 40°C (32 to 104°F). For operating
conditions, see
Table
This unit supplies power to the instrument and charges the
battery (if it is installed). All the instrument functions are
available with or without the battery.
(Section
Charge time (to full capacity)
|
6.5 hours
|
6.5 hours
|
13 hours (with a 500 mA
supply)
Battery duration
> 10 hours
> 6 hours
WARNING
10-1.
Prepare the instrument 2-3
Issue 1
2.6) or

Hide quick links: