Example Procedure: Source Dc Voltage (Ch1); Example Procedure: Measure Or Source Current With Internal 24V Loop Power - GE Druck DPI 620 User Manual

Advanced modular calibrator
Hide thumbs Also See for Druck DPI 620:
Table of Contents

Advertisement

3.2.6 Example procedure: Source DC voltage (CH1)

A
Source DC volts on
channel 1 (CH1)
Range: 0 to 12 V (DC)
Automation: Nudge
(Section
2.11.2)
B
Source DC millivolts on
channel 1 (CH1)
Range: 0 to 2000 mV (DC)
Automation: Nudge
(Section
2.11.2)

3.2.7 Example procedure: Measure or source current with internal 24V loop power

A
Measure current on
channel 2 (CH2) + internal
loop power (24 V)
Range: ±55 mA
B
Source current on
channel 2 (CH2) + internal
loop power (24 V)
Range: 0 to 24 mA
Automation: Nudge
(Section
2.11.2)
[EN] English - K0449
These examples (A and B) show Channel 1 (CH1) set up to
source a DC voltage.
1
3
1
3
1. Set the applicable software options; see
(Procedure
overview).
2. Complete the electrical connections.
3. To continue, set the applicable output value; see
Section
2.11.2.
These examples (A and B) show Channel 2 (CH2) set up to
measure or source a current with internal loop power (24 V).
1
1
3
2
2
Section 3.2.1
2
Maximum: 24 mA
2
Electrical and IDOS operations 3-5