2.10.3 Procedures to make Task selections

CH1:
CH2:
source
measure
Procedure Overview
2
1
TAP
A
B
C
Note: * Observed (Only
available as a source
option). Use this option to
make a manual record of
the readings on another
instrument; see
Section 2.11.3
[EN] English - K0449
When you first use the calibrator, there are default
measure
and source
CH1 settings: RTD source, RTD type is PT100, scale is °C;
Automation is Nudge; see
CH2 settings: Current measure, see
Use the Task menu to complete these procedures:
Set the calibrator functions you want to use on the display:
item
A
; see
Section
a. CH1: Channel 1 electrical function: You can choose one
option:
None
Current
Voltage
Millivolts
TC (°C/°F)
TC (mV)
Frequency
b. CH2: Channel 2 electrical function: You can choose one
option:
None
Current
Current (24V)
c. * Pressure function (P1): for the PV 62x pressure
stations, refer to user manual - K0457; for the MC 620
module carrier, see
d. * Pressure function (P2): the MC 620 module carrier can
use P1 and/or P2; see
e. * IDOS function: Universal pressure module; see
Chapter
3.
f. HART function: HART device communications; see
Chapter 8
If applicable, change the Units for the function: item
see
Section 2.10.5
functions on the display:
Chapter
3.
Chapter
2.10.4.
Resistance
RTD
CJ
AC Volts
AC mV
AC Probe
Observed*
Voltage
Millivolts
Observed*
Chapter
4.
Chapter
4.
Prepare the instrument 2-9
Issue 1
3.
;
B