Options; Awt 01 Temperature Sensor; Nurse Call - Dräger Savina 300 Technical Documentation Manual

Intensive care ventilator
Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

4.3

Options

4.3.1

AWT 01 temperature sensor

Prerequisites
The Savina 300 is connected to the mains power supply and to the „O2" com-
pressed gas supply and switched on.
Action
• Connect the AWT 01 temperature sensor to the Savina 300.
Action
• Perform a comparative measurement with a reference thermometer at
Test
The measured value of the reference thermometer matches the Savina 300
reading and is within a tolerance of 2 °C.
Entry
4.3.2

Nurse call

Prerequisites
The Savina 300 is connected to the mains power supply and to the „O2" com-
pressed gas supply and switched on.
Test set-up
Fig. 15
Item 1
1
3
4
5
Action
• Set „VC-CMV/VC-AC" mode.
• Set the following parameters:
– O2: 21%
– VT: 500 mL
– Ti : 2.0 s
– f: 12/min
• Press the „Alarms" softkey and set the following parameters:
– Paw: 100 mbar
• Press the „System configuration" softkey.
room temperature.
Test setup; central alarm
Designation
Pin 1 (no alarm)
Pin 3
Multimeter (resistance measurement)
Pin 5 (alarm)
Function and condition test

Options

[________OK]
43 / 50