Function and condition test
User-specific settings
4.4
User-specific settings
4.4.1
Input of user-specific settings
Action
• Re-enter all recorded user-specific settings from the „Device configuration"
Result
The user-specific settings have been re-entered into the Savina 300.
4.5
Final action
Prerequisites
– The test instructions have been performed as specified.
– All tests performed were passed successfully.
4.5.1
Device handover with test label
Action
• Attach a test label to the device.
• Supply the user/owner with a fully functioning device.
Entry
46 / 50
section (test item: Input of user-specific settings) in the Savina 300.
[________OK]
[________OK]