Rear panel
Preconditions
Removal
Savina 300 has been disconnected from the O2 supply and the mains
power supply.
The tubing system and units mounted on the Savina 300 have been
removed.
EGB conditions to protect against electrostatic discharge have been
established.
The filter cover has been removed, see section headed „Filter cover".
1. Pry the tab fuse
(Fig.
similar tool.
2. Remove the connectors and tubing listed below from the Savina 300, if
fitted:
Fig. 25
View of device rear; connections
Item
1
Connection (external battery)
2
Tab fuse
3
Mains connection
4
Connection (nurse call)
5
Connection (RS232 port)
6
Connection for LPO (option)
7
Connection (O2 gas supply)
Connector for potential equalization
Savina 300
Maintenance instructions
Disassembling/assembling the device
25/2) from its holder using a small screwdriver or
Designation
35