Dräger Savina 300 Technical Documentation Manual page 58

Intensive care ventilator
Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

Maintenance instructions
Replacing the external batteries
Fitting
58
10. Dispose of used batteries in accordance with local waste disposal regula-
tions.
Removal of the batteries is complete.
CAUTION
Risk of device failure due to damaged insulating strip!
If the insulating strip is damaged, a short-circuit may occur in the battery
pack and result in device failure.
Check the condition of the insulating strip before fitting the batteries.
1. Check that the insulating strip
necessary.
Fig. 56
Battery pack for external batteries; insulating strip
Savina 300
(Fig.
56/1) is not damaged; replace it as

Hide quick links: