Dräger Savina 300 Technical Documentation Manual page 55

Intensive care ventilator
Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

Removal
Preconditions
Procedure
The Savina 300 has been switched off and disconnected from the mains
power supply.
The connector of the DC battery cable has been disconnected from the
power supply unit of the Savina 300.
1. Actuate all locking brakes on the double-wheel castors.
2. Remove the screw
(Fig.
Fig. 51
Detail view of trolley; screw
3. Remove the screws and split washers
nect the electrical cable with integrated tab fuse (25 A/32 V)
Fig. 52
View of external batteries; screws, split washers and electrical
cable
4. Remove the screws and split washers
The batteries are fixed in the battery pack by a strap. The central screw holds
the strap in place.
Savina 300
Maintenance instructions
Replacing the external batteries
51/1), lift off the cover and place it aside.
(Fig.
52/1) (2 of each) and discon-
(Fig.
52/2) (2 of each).
(Fig.
52/3).
55