4.2.3
Emergency respiratory valve (D2)
Prerequisites
The Savina 300 is not connected to the mains power supply or to the „O2"
compressed gas supply. Savina 300 is switched off.
Action
• Prepare the following test setup:
Test set-up
Fig. 10
Item
1
2
3
4
5
6
7
8
9
10
11
Action
• Using the test pressure regulator, set a flow of 57 to 63 L/min.
Test
The measured pressure is in a range from -3 mbar to -6 mbar.
Entry
„Emergency respiratory valve" test setup
Designation
Test pressure regulator
Injector
Silicone tube
Connecting sleeve
ISO socket
Flowmeter
Adult ventilation tube (tube length 0.9 to 1.2 m)
Sealing plug
Y-piece, adult, Luer-Lock
Silicone tube
Pressure gauge
Function and condition test
Function tests
[________OK]
35 / 50