Emergency Respiratory Valve (D2) - Dräger Savina 300 Technical Documentation Manual

Intensive care ventilator
Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

4.2.3

Emergency respiratory valve (D2)

Prerequisites
The Savina 300 is not connected to the mains power supply or to the „O2"
compressed gas supply. Savina 300 is switched off.
Action
• Prepare the following test setup:
Test set-up
Fig. 10
Item
1
2
3
4
5
6
7
8
9
10
11
Action
• Using the test pressure regulator, set a flow of 57 to 63 L/min.
Test
The measured pressure is in a range from -3 mbar to -6 mbar.
Entry
„Emergency respiratory valve" test setup
Designation
Test pressure regulator
Injector
Silicone tube
Connecting sleeve
ISO socket
Flowmeter
Adult ventilation tube (tube length 0.9 to 1.2 m)
Sealing plug
Y-piece, adult, Luer-Lock
Silicone tube
Pressure gauge
Function and condition test
Function tests
[________OK]