Reading Out Customer-Specific Settings - Dräger Savina 300 Technical Documentation Manual

Intensive care ventilator
Hide thumbs Also See for Savina 300:
Table of Contents

Advertisement

Action
• Read off and note down the „Operating hours" reading.
Entry
Operating hours
Action
• Read off and note down the „Service hours" reading.
Entry
Service hours
1.1.4

Reading out customer-specific settings

Prerequisites
Savina 300 is connected to the mains power supply and is switched on.
Savina 300 is in „Standby" mode.
NOTE
After completion of the test procedure, re-enter the customer-specific set-
tings before handing the unit over to the customer/owner.
Action
• Press the „Ventilation settings" softkey.
• Select the „VC-CMV/VC-AC" ventilation mode and read off and note down
Entry
O2
Entry
VT
Entry
Ti
Entry
f
Entry
FlowAcc
Entry
PEEP
Action
• Select the „Advanced settings" submenu.
• Read and note down the following parameters from the display:
Entry
Sigh (ON or OFF)
Entry
Flow trigger (ON or OFF)
Entry
AutoFlow (ON or OFF)
Entry
int PEEP
Entry
Trigger
Action
• Press the „Alarms" softkey.
• Read and note down the following parameters from the display:
the following parameters from the display:
Device configuration
Savina 300
[________h]
[________h]
[________txt]
[________txt]
[________txt]
[________txt]
[________txt]
[________txt]
[________txt]
[________txt]
[________txt]
[________txt]
[________txt]