3.1.8

Optional multiple socket strip on trolley

Power fuse links
Action
• Compare the power fuse links with rated values on the multiple socket
Test
The power fuse-links match the specifications on the multiple socket strip.
Result
Protective conductor resistance
Action
• Check the protective conductor contacts of all sockets on the multiple
Test
The value of the protective earth conductor resistance must not exceed
0.3 ohms in each case.
Entry
Document the highest protective conductor resistance measurement in the
„Test Report".
Equivalent device leakage current
NOTE
The equivalent device leakage current test applies to the multiple-socket
outlet only!
NOTE
Always enter the initial value in a new „Test Report" document.
Test
The initial value must not exceed 100 μA.
Entry
Initial value
Test
The recurrent test value must not exceed 100 μA.
Entry
Recurrent test
strip.
socket strip with tester probe.
Electrical safety
Electrical safety to EN 62353
[________OK]
[________Ohm]
[________μA]
[________μA]
23 / 50