  • Page 1 Keysight Technologies MXA Signal Analyzer Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 to < MY/SG/US5606 Installation Note...
  • Page 2 The information contained in this document is subject to change without notice. Keysight Technologies makes no warranty of any kind with regard to this material, including but not limited to, the implied warranties of merchantability and fitness for a particular purpose.
  • Page 3 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Products Affected: N9020A, MXA Signal Analyzer Serial Numbers: ≥ MY53280000 to < MY56060000 ≥ SG53280000 to < SG56060000 ≥ US53280000 to < US56060000 To Be Performed By: (X) Keysight Service Center...
  • Page 4: Installation Kit Parts List

    Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Installation Kit Parts List Keysight Part Quantity Description Number A13 Front End Assembly N9020-60314 A15 Front End Controller w/ IF MUX N9020-60172 Adapter-Coaxial Straight Female-SMA to Female-SMA, 50 Ω 1250-1666 Washer, Lock, Internal Tooth, ¼-inch 2190-0067 Nut-Hex-Double-Chamfer ¼-36-THD .125-IN-THK Stainless Steel...
  • Page 5: Tools Required

    Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Tools Required — T-10 TORX Driver — T-20 TORX Driver — 5/16-inch torque wrench — ¼-inch open-end wrench — 9/16-inch nut driver — ¼-inch socket on 4-pound torque wrench — Scissors or knife —...
  • Page 6: Installation Procedure

    Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Installation Procedure Analyzer Information 1. Connect a power cord to the analyzer and turn on the analyzer. 2. After the analyzer has completed turning on, press System, Show, System. Make note of the following information from the Show System screen: Product Number: _____________ Serial Number: ______________...
  • Page 7: Update Instrument Software

    Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 9. Refer to the data in step 4 above. Verify that the Front End assembly has a HW ID of 70. 10.Refer to the data in step 3 step 4 above.
  • Page 8 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Licensing the New Options 1. Locate the Option Upgrade Entitlement Certificate (5964-5178) from the kit. 2. Redeem the Option Upgrade Entitlement Certificate by following the instructions on the Certificate. 3. After redeeming your Option Upgrade Entitlement Certificate you will receive an email with an attached License File.
  • Page 9 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 7. The signal analyzer will automatically consume the License File. (This may take a few minutes) When the License File is consumed the Keysight License Manager will display a “Successful License Installation”...
  • Page 10: Analyzer Disassembly

    Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Analyzer Disassembly If the instrument is placed on its face during any of the following procedures, be sure to use a soft surface or soft cloth to avoid damage to the front panel, keys, or input connector. If the analyzer has Option PRC, Portable Configuration, refer to the “Portable Instrument (Option PRC)”...
  • Page 11 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Standard Instrument (Benchtop Configuration) 1. Disconnect the instrument from ac power. 2. Refer to Figure 3. Using the T-20 driver, remove the four screws (two on each side) (1) that attach the handle strap (2) on each side of the instrument.
  • Page 12 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Portable Instrument (Option PRC) Make sure any adapters on the front panel are removed. 1. Disconnect the instrument from ac power. 2. Refer to Figure 4. Using the T-20 driver, remove the four screws (two on each side) (1) that hold the bail handle (2) to the front frame.
  • Page 13 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 4. Refer to Figure 5. Using the T-20 driver, remove the four screws including washers (1) that hold the rear bumpers (2) in place. 5. Pull the instrument cover (3) off towards the rear of the instrument. Figure 5 Option PRC Instrument Outer Case Removal Installation Note N9020-90230...
  • Page 14 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Front Frame Assembly Removal 1. Refer to Figure 6. Using the T-10 driver, remove the eight screws (1), four on each side, to detach the front frame from the chassis. Figure 6 Front Frame Removal Installation Note N9020-90230...
  • Page 15 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 2. Refer to Figure 7. Pull the front frame carefully away from the chassis. Remove the ribbon cable W1 from the A8 Motherboard. W1 may have locking springs on each side. Depress the spring on each side of the connector to remove from the motherboard.
  • Page 16 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Rear Panel Removal 1. Refer to Figure 8. Using the T-10 driver, remove the screws (2) attaching the rear panel (1) to the chassis. 2. Use a 9/16-inch socket wrench to remove the nut securing the EXT REF IN connector from the rear panel.
  • Page 17 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Top Brace Removal 1. Refer to Figure 9. To remove the top brace (1), use the T-10 driver to remove the eight panhead screws (3) (0515-0372), four on each side, attaching the braces to the chassis. Also remove the six flathead screws (4) (0515-1227) attaching the top brace to the boards.
  • Page 18 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Replace A13 Front End Assembly Refer to Analyzer Information at the beginning of the Installation Procedure. Only perform this procedure if the Front End has a Hw ID of 70 and the part number of the Front End is either E6607-61002 or E6607-61004.
  • Page 19 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Figure 10 RF Front End Assembly Removal 5. Locate the A13 RF Front End Assembly in this kit. Refer to Figure 10. Place the RF Front End Assembly into the chassis. Replace the four screws (2). Using the T-10 driver, torque to 9 inch-pounds in the sequence shown, starting with #1.
  • Page 20 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Replace Front End Control Assembly Refer to Analyzer Information at the beginning of the Installation Procedure. Only perform this procedure if the Front End Controller has a Hw ID of 71. 1.
  • Page 21 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 3. Refer to Figure 12. Install the Enhanced Front End Control assembly included in this kit into slot 6 in the chassis securing with the ejector. 4. Reattach the ribbon cables W16, W17, and W18 to the Front End Control assembly (1). 5.
  • Page 22 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 6. Refer to Figure 13. Disconnect cables W12 and W13 from the bottom of the DIF assembly. Figure 13 Disconnect W12 and W13 Installation Note N9020-90230...
  • Page 23 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 7. Refer to Figure 14. Pull the AIF/DIF assembly part way out of the chassis. Disconnect cable W15 from the analyzer. 8. Completely remove W15 from the analyzer. Figure 14 AIF/DIF Assembly Partial Removal 9.
  • Page 24 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 12.Refer to Figure 15. Reconnect W12 to A3J14. Reconnect W13 to A3J15. Figure 15 Reconnect W12 and W13 13.Locate W39 in the Opt EXM/HL6 Cable Kit. This cable will have part number 8121-2028 and will be labeled “1100”...
  • Page 25 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Add Cables to A13 Front End and A15 Front End Controller 1. Locate the flexible coax assembly in the Opt EXM/HL6 Cable Kit with Markers that is labeled “8120-2027”. The ends should be labeled “903” and “13”. This is W26. 2.
  • Page 26 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 5. Refer to Figure 17. Dress the coaxial cables and semi-rigid cables neatly and tie together using cable ties (1400-0249) at the locations indicated. Figure 17 Adding Cable Ties near A13 and A15 Assemblies Installation Note N9020-90230...
  • Page 27: Add Ext Mixer Connector And Cable To Front Frame Assembly

    Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Add EXT MIXER Connector and Cable to Front Frame Assembly 1. Locate the SMA female to SMA female connector (1250-1666), ¼” lock washer (2190-0067), and ¼” hex nut (2950-0223) in the kit. If the SMA connector includes hardware, discard that hardware and use the 2190-0067 lock washer and...
  • Page 28 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Figure 19 Connecting W27 to Ext Mixer Connector Figure 20 Proper Orientation of W27 Installation Note N9020-90230...
  • Page 29 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Front Frame Replacement 1. Refer to Figure 21. Reattach the ribbon cable W1. Ensure the locking tabs are engaged. 2. Reattach the ribbon cable W24 for instruments with Option BBA. Figure 21 Front Panel Cable Installation Note N9020-90230...
  • Page 30 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 3. Refer to Figure 22. Carefully position the Front Frame Assembly onto the chassis. Ensure no cables are crushed. Replace the eight screws (1), four on each side of the chassis. Torque to 9 inch-pounds.
  • Page 31 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 4. Refer to Figure 23. When the front panel is installed, cables W27 and W28 typically do not align. This is OK since the cables are flexible. 5. Refer to Figure 24.
  • Page 32 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 6. Refer to Figure 25. Use a ¼” open-end wrench to prevent the SMA female connector on W28 from twisting. Use a 5/16” torque wrench to torque the nut on W27 to 10 inch-pounds. Figure 25 Torque Cable W28 onto W27 Installation Note N9020-90230...
  • Page 33 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Reassembly 1. Refer to Figure 26. To replace the top brace (1), place it in the correct position and attach the appropriate screws. Torque to 9 inch-pounds. Figure 26 Top Brace Replacement 2.
  • Page 34 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 5. Refer to Figure 27. Place the rear panel (1) into position in the chassis. Replace the screws (2) to attach the rear panel to the chassis. Torque to 9 inch-pounds. Figure 27 Rear Panel Replacement 6.
  • Page 35 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Figure 28 Add Warning Label to Top Brace Installation Note N9020-90230...
  • Page 36 Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Final Installation for Standard Instruments (Benchtop Configuration) 1. Refer to Figure 3. Carefully slide the instrument cover back onto the instrument from the rear of the analyzer, making sure not to damage any internal cables. The seam on the cover should be on the bottom of the instrument.
  • Page 37: Verify Hardware Installation

    Option HL6, External Mixing Upgrade Kit, Serial Prefix ≥ MY/SG/US5328 Verify Hardware Installation 1. Verify the spectrum analyzer application loads and sweeps as expected. 2. Press System, Show, Hardware on the analyzer and verify that the Front End and Front End Controller assemblies identify themselves as: —...
  • Page 38: Adjustments Required

    For assistance, contact your nearest Keysight Technologies Sales and Service Office. To find your local Keysight office access the following URL, or if in the United States, call the following telephone number: http://www.keysight.com/find/assist...