Safety and handling

Bear in mind the following safety considerations when performing
maintenance procedures.
Warning:
biohazardous materials, always wear chemical-resistant,
powder-free gloves at all times while handling the components.
Warning:
removal of contaminating deposits using flammable or caustic
agents. You must be aware of the risks, and take suitable
precautions.
Warning:
standby before commencing any maintenance.
Warning:
these components. The probe and source may be hot.

Removing O-rings and seals

When performing certain maintenance procedures, you will need to remove
O-rings or seals from instrument components. An O-ring removal kit is
provided with the instrument. You must dispose of all O-rings and seals that
have been removed; do not re-use old O-rings or seals on the instrument.
O-ring removal kit
Tool 1
To avoid self contamination with toxic and
Many of the procedures in this chapter involve the
To avoid electric shock, ensure that the instrument is in
To avoid burn injuries, take great care while working with