CARESCAPE Monitor B850

7.9 Setting power frequency

WARNING
Incorrect power line frequency setting could adversely affect ECG and EEG
processing.
1.
Log in to Webmin.
2.
Select Configuration > Power Frequency.
3.
In the Power Frequency window, select the applicable power line frequency.
4.
Select Save.
The power frequency configuration takes effect immediately.

7.10 Selecting language and locale

Select the language used in clinical user interface and select the keyboard locale setting for the
alphanumeric keyboard and the barcode reader.
1.
Log in to Webmin.
2.
Select Configuration >Language.
3.
In the Language window, select the patient monitor language and keyboard language:
a.
b.
The language takes effect after the patient monitor is restarted. The keyboard locale takes
effect immediately after it is submitted.

7.11 Selecting national requirements

Activate France specific defaults for the ECG HR adjustment range and the reminder beep
behavior.
1.
Log in to Webmin.
2.
Select Configuration > National Requirements.
3.
In the National Requirement window, select the applicable option:
Value
None
France
4.
Select Save.
The national requirements changes take effect immediately.

7.12 Configuring modules

You can configure some acquisition module specific settings. The settings are saved to the
permanent memory of the related acquisition module and the settings travel with the module
from one patient monitor to another.
These settings are pre-configured at factory, except the Assets Settings. You may need to
re-configure them after corrective maintenance, or for administration purposes.
84
2106778-001
Select the patient monitor language from the drop-down list and select Save.
Select the keyboard locale from the drop-down list and select Save.
Description
Select the normal defaults.
Enable the following country specific monitoring:
-
Heart Rate high alarm limit maximum 230.
-
Reminder beep will sound every 2 minutes when alarms have
been silenced permanently.