Configuring Adt Server Settings; Setting Time And Date; Setting Unit And Bed Name - GE CARESCAPE B850 Service Manual

Monitor
Hide thumbs Also See for CARESCAPE B850:
Table of Contents

Advertisement

CARESCAPE Monitor B850
The S/5 virtual ID change will be applied immediately. All other network configurations will be
saved and become active when the patient monitor is restarted

7.2.4 Configuring ADT server settings

1.
Log in to Webmin.
2.
Select Configuration > Network > ADT settings.
3.
In the ADT settings window, select Enable to enable the Visit Number Query.
4.
Enter the ADT IX IP address.
NOTE: See section
addresses that must not be used as the ADT IX IP address.
5.
Select Save.
To disable the Visit Number Query, select Disable in the ADT settings window.

7.3 Setting time and date

NOTE: In the S/5 Network, the patient monitor's date and time is automatically synchronized
with the date and time of the iCentral it is connected to. You can set the date and time locally
only if the patient monitor is not connected to the S/5 Network and the patient monitor is in a
discharged state.
CAUTION
NETWORK DEVICE TIME SYNCHRONIZATION — When adding a new device to
the CARESCAPE Network, the existing devices on the CARESCAPE Network will
synchronize to the new device's time. To prevent potential time
synchronization issues, you should set the new device's time to be as close as
possible to the time (within one minute or less) used by the existing GE devices
on the CARESCAPE Network.
1.
Log in to Webmin.
2.
Select Configuration > Time.
3.
In the Time Configuration window, in Configure Date and Time, update the following
fields as needed:
4.
In the Time Configuration window, in Configure Time Zone, select the appropriate UTC
Offset, if applicable.
NOTE: The time zone configuration only applies to communication with Citrix server.
5.
Select Save.
The manual time configuration takes effect immediately.

7.4 Setting unit and bed name

Configure the care unit name and bed name for patient monitors that are configured to
connect to the MC Network.
72
2106778-001
7.2.2 Selecting and configuring CARESCAPE Network
Date
Month
Year
Hour:Minute
AM/PM
12/24 Hrs
for a list of reserved

Advertisement

Table of Contents
loading

Table of Contents