The pending software activation is cancelled and the patient monitor continues running the
current software version.

7.20.4 Erasing an inactive software version

You can erase an inactive host software version from the patient monitor. Erasing can prevent
the activation of a wrong software version by mistake.
1.
Log in to Webmin.
2.
Select Configuration > Software Management.
The Software Management window opens.
3.
Select the software that you want to erase and select Next.
The selected software activation window appears.
4.
Select Erase to erase the inactive software.