Installation checkout

The purpose of the installation checkout procedure is to ensure that the system is properly
installed and configured for use.
The installation check covers the CARESCAPE B850 monitoring system including the following
devices:
CARESCAPE B850 host
Displays
F5/F7 Frames
E-modules
CARESCAPE PDM
Service personnel shall perform the following checkout procedure for the monitoring system
after the hardware installation and platform configuration is completed:

8.1. Visual inspection

8.2. Electrical safety tests
NOTE: The manufacturer has performed the electrical safety test for the patient monitor
and acquisition modules during final inspection. You do not have to perform the electrical
safety tests during installation checkout, if there is less than 12 months since the patient
monitor was manufactured. Check the date of manufacture of the device from the device
label (see section
8.3. Functional check
NOTE: Refer to chapter
procedure after corrective and planned maintenance.
8.1 Visual inspection
Perform the following visual inspection to the installed monitoring system:
Carefully inspect the patient monitor and the connected peripheral devices for any
damage.
Verify that the patient monitor and the connected peripheral devices are properly
mounted with specified mounting solutions.
Verify that the cables between the patient monitor and the connected peripheral devices
are intact, properly connected and secured to the right connectors.
Verify that the power cord and USB cables are properly secured with the supplied
retaining clips.
Verify that all the network cables, both MC and IX Network, are intact and properly
connected to the right connectors.
Verify that the modules are properly connected and locked.
The cleaning precautions, requirements, procedures, and recommended cleaning solutions for
the patient monitor are described in the patient monitor's user's manual. For details about
cleaning, disinfecting and sterilizing the accessories, see the instructions for use in the
accessory package.
3.5.2. Equipment
identification).
10. Maintenance and checkout
Installation checkout
to see the recommended checkout
105
2106778-001

Advertisement

Table of Contents
loading

Table of Contents