S/5 Network
Problem
Bedside monitor doesn't
register to the Monitor Network
- all bedside monitors affected.
Bedside monitor doesn't
register to the Monitor Network
– only one patient monitor or
some of the bedside monitors
affected.
Possible cause
The problem is related to the
iCentral, or to the cabling
between the iCentral and the
hub/switch.
The problem is related to the
bedside monitor, or to the
cabling between the bedside
monitor and the hub/switch.
The problem is related to
network setup: the bedside
monitor(s) is/are missing from
the network setup.
The problem is related to view
setup: the bedside monitor(s)
is/are missing from the view
setup.
Troubleshooting
Recommended action
Proceed to the iCentral
troubleshooting.
Check the cabling between
the iCentral and the
hub/switch.
Check the patient monitor(s)'
network configuration.
Check the cabling between
the bedside monitors and the
hub.
Check the iCentral network
setup.
Check the iCentral view setup.
151
2106778-001