Incorrect system time
11.5.3
Problem
System time is incorrect when
patient monitor is not
connected to network.
System time is incorrect when
patient monitor is connected to
MC Network.
System time is incorrect when
patient monitor is connected to
S/5 network.
11.5.4 License issues
Problem
Unable to perform a function or
a feature is not available.
Possible cause
CPU timekeeper battery empty
Time not configured properly
Network device time
synchronization error.
iCentral time incorrect.
Possible cause
-
A license has not been
purchased for the feature.
-
The trial license has expired
for the feature.
-
The license is not installed
properly.
Recommended action
Replace CPU timekeeper battery.
Configure date and time, see 7.2.4.
When adding a new device to the
CARESCAPE Network, the existing devices
on the CARESCAPE Network will
synchronize to the new device's time. To
prevent potential time synchronization
issues, you should set the new device's
time to be as close as possible to the time
(within one minute or less) used by the
existing GE devices on the CARESCAPE
Network.
Configure iCentral time master.
Recommended action
See License management chapter.