GE CARESCAPE B850 Service Manual page 118

Monitor
Hide thumbs Also See for CARESCAPE B850:
Table of Contents

Advertisement

CARESCAPE Monitor B850
NOTE: *The measuring device (MD) represents the network and voltage measuring instrument
and its frequency characters according to IEC 60601-1.
1.
Configure the safety analyzer as follows (NC):
2.
Power on the device under test.
3.
Read and record the current leakage indicated on the safety tester.
4.
Configure the safety analyzer as follows (SFC):
5.
Read and record the current leakage indicated on the safety tester.
6.
Configure the safety analyzer as follows (SFC):
7.
Read and record the current leakage indicated on the safety tester.
8.
Configure the safety analyzer as follows (SFC):
110
2106778-001
Polarity: NORMAL
Neutral: CLOSED
Earth (GND): CLOSED
Polarity: NORMAL
Neutral: OPEN
Earth (GND): CLOSED
Polarity: NORMAL
Neutral: CLOSED
Earth (GND): OPEN
Polarity: REVERSED
Neutral: CLOSED
Earth (GND): OPEN