Required Tools; Before Disassembly - GE CARESCAPE B850 Service Manual

Monitor
Hide thumbs Also See for CARESCAPE B850:
Table of Contents

Advertisement

CARESCAPE Monitor B850
GE recommends using the new fasteners (screws, washers, etc.) in the FRU kits rather
than reusing the old fasteners. Some fasteners are not intended to be re-used more than
three times.
Use only new screws when attaching parts into light metal parts. Before fastening a
screw, turn it counterclockwise until it drops into an existing thread pattern.

12.1.3 Required tools

WARNING
Due to possible high voltage present, use an insulated screwdriver at all
times.

12.1.4 Before disassembly

WARNING
PATIENT MONITORING INTERRUPTION — Make sure a patient is not being
monitored while servicing the equipment.
WARNING
DISCONNECTION FROM MAINS - When disconnecting the device from the
power line, remove the plug from the wall outlet first. Then you may
disconnect the power cord from the device. If you do not observe this
sequence, there is a risk of coming into contact with line voltage by
inserting metal objects, such as the pins of leadwires, into the sockets of the
power cord by mistake.
WARNING
SAFETY GROUND - Remove power cord from the mains source by grasping
the plug. Do not pull on the cable.
WARNING
ELECTRIC SHOCK - Always unplug the grounded cables when not in use.
Leaving them connected could result in an electric shock from the ground
contact in the other end.
1.
Turn the monitor off from the rear
2.
Detach the retaining clips for the power cord and USB cables.
NOTE: Remember to reattach the retaining clips for the power cord and USB cables after
reassembly.
154
2106778-001
-
insulated Torx T10 screwdriver
-
insulated crosshead screwdrivers (small and
medium)
-
antistatic ESD wristband
-
spanner size 5.5 mm
On/off
switch.