activation process with a delay during an active patient case, but the new settings will activate
only after the next case reset / patient discharged.
1.
Log in to Webmin.
2.
Select Configuration > Settings.
3.
In the Sub-Modules for Settings menu, select Activate.
4.
In the Activate Settings window, select the settings that you want to activate:
Select Cloned settings (clinical and platform) to activate both the clinical and
platform settings.
Select Cloned clinical settings to activate only loaded clinical settings.
Select Cloned platform settings to activate only loaded platform settings.
NOTE: Activation of the cloned clinical or platform settings will not affect the following
settings in the target patient monitor:
Network
Unit and Bed Name
Host Asset Settings
Licenses
Remember to configure the unaffected settings manually in the target patient monitor.
Select Factory default settings to restore all platform and clinical settings back to
the factory defaults.
Select US default clinical settings to restore all clinical and platform settings back
to the US specific factory defaults.
NOTE: Activation of the factory defaults and US factory defaults will leave the following
settings in the target patient monitor unaffected:
- Host Asset Settings
- Licenses
5.
In the Setting Activation, select whether you want the new settings to take effect
immediately or after the next case end / discharge:
Configuration
91
2106778-001