Iec 60601-1 Classifications - GE CARESCAPE B850 Service Manual

Monitor
Hide thumbs Also See for CARESCAPE B850:
Table of Contents

Advertisement

CARESCAPE Monitor B850
CPU unit
1.
Power on/off switch
2.
Power inlet connector
3.
Equipotential connector
4.
Four USB ports for connecting the touchscreen display, remote control, keyboard, mouse,
and barcode reader
5.
Two Ethernet network connectors for connecting the MC and IX networks. The MC
Network establishes communication and allows patient data to be sent to a CARESCAPE
Central Station. The IX Network provides access for example to the MUSE server, Citrix
server, and IX printers.
6.
Two ePort connectors for connecting the PDM and E-module Frame.
7.
Two serial RS232 connectors
8.
DVI-D 2 connector for one digital display
9.
Optional 3rd video display connector, DVI-I 3 supports a digital display and a cloned
analog display (iPanel application only).
10. DVI-I 1 connector for a digital display and a cloned analog display.

3.4 IEC 60601-1 classifications

Type of protection against electric shock: Class I.
Degree of protection against electrical shock: applied parts are marked with a symbol
indicating degree of protection.
Degree of safety of application in the presence of flammable anesthetic mixture with air
or with oxygen or nitrous oxide: Not suitable.
WARNING
EXPLOSION. Do not use this equipment in the presence of flammable
anesthetics, vapors or liquids.
Mode of operation: Continuous.
Method(s) of sterilization or disinfection recommended by the manufacturer: see the
user's manual.
IEC 60601-1-2
The system complies with IEC 60601-1-2:2007-03.
In accordance with IEC 60601-1-2, modules marked with the ESD warning symbol (IEC
60417-5134) require user training in ESD awareness and prevention as follows:
20
2106778-001