System Installation

Initializing the Banded Millimeter-wave System

Initializing the Banded Millimeter-wave System
After the PNA has been received from the factory, you need to change your
PNA over to "Millimeter Config":
1. Ensure that PNA and N5292A test set are powered on.
2. On the PNA: press Setup > External Hardware > Millimeter Config
3. The PNA application will close and reopen.
4. Verify the PNA is sweeping across the band that was specified in
For N5293A/5A, the banded millimeter-wave system should now initialize and
recognize the test set modules. The port 1 module LED should be blue
(addressed by the PNA and active) and the other test port(s) should be green
(no bias and on standby). See also,
Modules" on page
Keysight Self-Configured System Installation Guide
"OML/VDI and N5293A/5A Test Heads – Banded Configuration Calibration
(Only)" on page
3-21.
3-33.
"Troubleshooting the OML/VDI MiIlimeter
3-29