Front Panel Cabling - Keysight PNA Series Operation And Installation Manual

2-port and 4-port microwave network analyzer system
Hide thumbs Also See for PNA Series:
Table of Contents

Advertisement

System Installation
PNA, Controller, and Test Head Module Interconnections

Front Panel Cabling

The front-panel connections between the millimeter head controller and a test
head module are shown in
Figure 3-12 on page
each test head module in the system. This cabling is duplicated for each test
head module in the system.
Figure 3-11
Example of Cable Connections for N5293A/5A Four Test Heads Module
Millimeter-wave System Cable Connections
This section describes a typical 4-port cable connection process on the
millimeter-wave system. 2-port cable connections are similar.
Do not connect the LFE semi-rigid cables. Instead connect 50 ohm
terminations (1810-0118). Refer to
4-port connections are shown, but 2-port cable connections are similar,
except where indicated.
This section assumes you have connected the rear panel connectors per
"Rear Panel Cabling" on page
1. Using the interconnect cables, connect all of the PNA's SOURCE OUT to
2. Install connections for each head. Refer to next the section,
Keysight Self-Configured System Installation Guide
17, and
the test set's Source In. Refer to
Connections" on page
Figure 3-11
and for OML/VDI heads, refer to Figure
Figure 3-13 on page
Figure 3-7 on page
3-14.
Figure 3-7 on page
3-16.
18. This cabling is similar for
12.
12.
"Test Head
3-15