Brightness Of Lc Display; Calibrating The Ph Meter; Automatic Buffer Detection - ProMinent DULCOMETER DMT Series Operating Instructions Manual

On-site measurement transducer measured variables ph/redox/temperature
Hide thumbs Also See for DULCOMETER DMT Series:
Table of Contents

Advertisement

8.1.2
You can alter the brightness of the LC display in the permanent display:
8.2
The pH probe condition is a key factor in the quality of the measurement. The
pH probe must therefore be re-calibrated at regular intervals with the aid of
buffer solutions. The DMT incorporates automatic buffer detection.
8.2.1
The tables for the following two buffer solution sets are contained in the
program memory.
Temp.
We recommend that you use the buffer solutions obtainable from ProMinent
for calibration purposes.
Select a buffer set which is higher than the general setting (see chapter 7.1.7)
If V = automatic adjustment has been selected the corrected values are
saved and used for buffer detection during the subsequent calibration
process.
On delivery the program memory contains the ProMinent buffer table as the
V-table.
ProMinent
®

Brightness of LC display

Increase brightness using the down arrow key
arrow key
.

Calibrating the pH meter

Automatic buffer detection

ProMinent buffer table
°
C
pH
0
4.05
7.13
5
4.04
7.07
10
4.02
7.05
15
4.01
7.02
20
4.00
7.00
25
4.00
6.98
30
4.00
6.98
40
4.00
6.97
50
4.00
6.96
60
4.00
6.97
70
4.01
6.98
80
4.02
6.99
Operating the DMT
or reduce using the up
DIN 19266 buffer table
10.26
10.17
4.00
10.11
4.00
10.05
4.00
10.00
4.00
9.94
4.01
9.90
4.01
9.82
4.04
9.75
4.06
9.68
4.09
9.62
4.13
9.55
4.16
pH
6.95
10.25
6.92
10.18
6.90
10.12
6.88
10.07
6.86
10.01
6.85
9.97
6.84
9.89
6.83
9.83
6.84
6.85
6.86