Handling and Transportation of the System
Storage of the System and Accessories
CUSA
Clarity Ultrasonic Surgical Aspirator System Operator's Manual
®
If you need to move/relocate the CUSA Clarity Ultrasonic Surgical
Aspirator System within the environment of the Healthcare Facility,
note the guidelines below:
Before moving the console:
•
Power off the system.
•
Unplug the power cord from the wall receptacle and secure it.
•
Disconnect the handpiece.
•
Disconnect and remove the irrigation fluid bag.
•
Lower the IV pole.
•
Secure any loose objects. Make sure there are no loose objects
placed on the top of the console.
•
Secure the footswitch and cord.
When moving the console (if attached to a cart):
•
Push the cart using the handle.
•
Do not run while pushing the console and cart.
•
Use an elevator to move the console between floors of a building.
Never use a stairwell.
•
Do not attempt to lift the console while attached to the cart.
CAUTION
To move the console up or down a ramp, use two or more people.
Console
•
Make sure the system is clean. Store the system in a low traffic area
that is free of dirt, blood, water, and other contaminants.
•
Contamination guard may be stored with console if it is not expired.
•
Store the system at an ambient temperature between -20°C
(-4°F) and 60°C (154°F).
Console with Cart
•
Wrap footswitch cord around cart and store footswitch in cart basket.
•
Wrap power cord around cart.
Handpiece Maintenance