Alarm Descriptions

Extracting Log Files

Alarm Descriptions
13-2
1. From the Main Screen, press the Settings button
2. Press System Information.
3. Attach a USB drive to the USB connector on the rear of the console.
NOTE
The external USB drive should be a FAT, FAT32, exFAT or NTFS
formatted drive. Other file formats are not supported.
4. Press the Extract Log button.
NOTE
When the extraction is complete, a message to remove the USB drive
displays.
5. Remove the USB drive.
Condition
Low Aspiration
Handpiece Failure
Footswitch Failure
Console Overheating
CUSA
Cause
The console has
measured the
aspiration level to be
too low to be
functioning properly.
A fault has occurred
reading or writing to
the handpiece
memory chip.
A footswitch wiring or
connection fault has
been detected.
A high temperature
has been measured
within the console.
The console will
automatically shut
off.
Clarity Ultrasonic Surgical Aspirator System Operator's Manual
®
Recommendation
• Check tubing
connections and
routing
• Replace tubing set
if problem persists
• Disconnect and
reconnect
handpiece
• Replace handpiece
if problem persists
• Disconnect and
reconnect
footswitch
• Replace footswitch
if problem persists
Contact Integra
Service
.