Integra CUSA Clarity Operator's Manual page 92

Ultrasonic surgical aspirator system
Table of Contents

Advertisement

CUSA
Clarity Ultrasonic Surgical Aspirator System Operator's Manual
®
WARNING
Before use, sterilize the torque base in the sterilization tray with the
handpiece.
CAUTION
To avoid product damage, always use the torque base to hold the
handpiece while using the torque wrench to tighten or loosen the tip.
Avoid over torqueing the tip.
6. Rotate the torque wrench clockwise until it clicks twice.
Remove the handpiece from the torque base.
7. Slide the larger end of the nosecone over the tip and slide it down over
the neck of the handpiece making sure the white dots are aligned.
8. Rotate the nosecone clockwise until it locks into place and the white
dot on the nosecone aligns with the while dot on the body of the
handpiece.
Make sure the white dots are aligned on the nosecone and handpiece.
Changing a Tip
10-7