Warnings and Cautions
1-8
CAUTION
To avoid product damage, always use the torque base to hold the
handpiece while using the torque wrench to tighten or loosen the tip.
Avoid over torqueing the tip.
CAUTION
Do not use sterilization methods involving temperatures in excess of
134ºC as this may damage the handpiece.
CAUTION
Product damage may result if you do not follow the instructions
described in the cleaning section.
CAUTION
To move the console up or down a ramp, use two or more people.
CAUTION
High frequency surgical equipment can affect the function of medical
electrical devices such as ultrasonic aspirators. When using monopolar
high frequency surgical equipment simultaneously with the CUSA
Clarity system, contact between the activated monopolar electrosurgery
instrument and the CUSA Clarity surgical tip could affect system
functionality. Avoid contact between the CUSA Clarity surgical tip
and an activated monopolar surgical instrument.
CUSA
®
Clarity Ultrasonic Surgical Aspirator System Operator's Manual